Fretting corrosion.pdf

(624 KB) Pobierz
PII: S0301-679X(99)00099-7
Tribology International 32 (1999) 587–596
www.elsevier.com/locate/triboint
Fretting corrosion of materials for orthopaedic implants: a study of
a metal/polymer contact in an artificial physiological medium
Bruno Tritschler * , Bernard Forest, Jean Rieu
D´partement M´canique des Biomat´riaux et Traitements de Surface, Ecole Nationale Sup´rieure des Mines de Saint-Etienne, 158, Cours
Fauriel, 42023 Saint-Etienne Cedex 2, France
Received 20 May 1999; received in revised form 15 November 1999; accepted 16 November 1999
Abstract
The fretting corrosion behaviour of a 316L SS flat against a PMMA counterface has been investigated in an artificial physiological
medium. A specific device has been used to visualize the in situ degradation at the contact interface. Simultaneous analysis of the
coefficient of friction and free corrosion potential has shown four distinct stages during fretting experiments. An energy-oriented
approach to the fretting process was conducted in tandem with measurement of wear. This method has shown a linear progression
in the wear volume of the samples as a function of the interfacial energy dissipated during fretting. The presence of chlorides
contributes to a considerable acceleration of the degradation of the stainless steel surface. This process was explained by a mechanism
related to crevice corrosion activated by friction.
Ó
2000 Elsevier Science Ltd. All rights reserved.
Keywords: Fretting corrosion; Wear; 316L SS; Orthopaedic implants
1. Introduction
trochemical reactions with the environment predominate,
are known as fretting corrosion [8,9].
The biomedical field is not immune to these phenom-
ena. Thus, the mechanical devices used in orthopaedic
surgery are subject to fretting [10,11]. These mech-
anisms have been used for many years in assemblies
such as screw/plate connections [12,13]. In association
with this phenomenon, problems of fretting corrosion in
total hip joint prostheses have been observed for several
years [14–16]. The most frequently employed method
for fixation is by cementing the femoral stem into the
bone using a polymer cement, poly(methylmethacrylate)
(PMMA). Fig. 1 shows the mode of implantation of
such prostheses.
Three groups of materials are widely used at present:
316L austenitic stainless steels, cobalt–chromium alloys
and Ti–6Al–4V titanium alloys, much used since the
1970s because of their good mechanical properties com-
bined with high biocompatibility [17].
Small movements develop either immediately on
implantation or after a few years of service. These small
amplitude movements are all the more dangerous for
occurring in an environment rich in chlorides. Instances
of loosening of the prosthesis can be explained in part
by the great difference in elastic modulus between the
When two surfaces in contact are subjected to an
oscillatory movement of low amplitude, surface deterio-
ration generally appears [1–4]. The phenomena of wear
induced by friction over small displacements, often
designated by the term fretting , can develop in any
mechanical device subjected to alternating movements,
such as vibrations or oscillations. Examples include
bolted assemblies, cables [5], rotors, riveted joints or
electrical contacts.
Damage by fretting involves all kinds of materials:
metals, ceramics [6] or polymers [7]. The most important
consequence is a drastic reduction in the lifetime of
mechanical systems, due to jamming of the parts in con-
tact or, in contrast, the appearance of play.
Environmental effects are of overriding importance in
the phenomena of fretting. Some corrosive media, such
as chlorides, contribute to a considerable acceleration of
wear. The processes involved, in which chemical or elec-
* Corresponding author. Tel.:
+33-4-77-42-02-51; fax:
+33-4-77-
42-00-00.
E-mail address: tritschl@sms.emse.fr (B. Tritschler).
0301-679X/99/$ - see front matter
Ó
2000 Elsevier Science Ltd. All rights reserved.
PII: S 0301-679X(99)00099-7
4970082.004.png 4970082.005.png 4970082.006.png
588
B. Tritschler et al. / Tribology International 32 (1999) 587–596
a UHMWPE hip prosthesis cup was fretted against a
316L SS sphere in Ringer’s solution, in a configuration
similar to the in vivo relationship between the spherical
head and the acetabular cup, a phenomenon of crevice
corrosion activated by the friction developed at the sur-
face of the metal, and considerable abrasion of the poly-
mer was observed.
The purpose of the present work was to study the
phenomena of wear by fretting corrosion of a metal,
316L stainless steel, in contact with a polymer,
poly(methylmethacrylate) (PMMA). The tests, per-
formed in an artificial physiological medium using a
device for in situ visualization of the contact surface,
led to the development of a qualitative and quantitative
approach to fretting corrosion.
Fig. 1. Diagram showing cemented implantation of a total hip joint
prosthesis.
2. Materials and methods
polymer and the metal femoral stem, and in part by
imperfect initial fixation resulting from the shape of the
prosthesis [18]. Fretting corrosion then causes the forma-
tion of debris in the form of a mixture of particles of
metallic oxides and polymer. The direct consequences
are inflammatory reactions of the surrounding bony
tissues and the necessity for a second surgery. Examples
are metallosis cases which are due not only to the cor-
rosion of the prosthesis in service, but more likely to the
emission and the effects of wear particles produced by
fretting [14]. The analysis of prostheses which have been
removed does not always reveal the actual reasons for
their failure. It was therefore necessary to develop test
methods to study the fretting behaviour of materials used
for implants. One solution consists in reproducing as
faithfully as possible the functioning of a hip joint pros-
thesis by means of walk simulators. Such devices, based
on the average behaviour of the patients, are difficult to
use and costly to create. Thus the need to develop tech-
niques of in vitro investigation led to the design of
devices capable of producing wear faces similar to those
observed on explanted prostheses, by means of simple,
rapid and reproducible tests. Many previous studies have
indeed examined the problems of deterioration of ortho-
paedic implants in general: experimental techniques have
been created to reproduce it and surface treatments
devised to remedy it [19–22]. Confinement appears to
be an essential factor in the deterioration. In the case of
large amplitude displacements, to which the term “fret-
ting” cannot be applied, the degree of deterioration is
observed to vary according to the degree of confinement.
Earlier studies were carried out on a pin made of
UHMWPE and a disc of 316L SS in an artificial physio-
logical solution. These showed almost no damage to the
surface of the metal [21], while slight polishing of the
surface of the polymer had occurred. In contrast, when
Cylindrical polymer specimens were rubbed against
stainless steel flats. Samples of PMMA were taken from
cylindrical bars of 20 mm diameter. They took the form
of blocks with a rectangular base of 9 mm
´
15 mm, with
10 mm. The PMMA
used was optically transparent (Altuglass
=
). The plane
surface opposite the contact interface was polished to
enable observation of the polymer/metal interface
through the transparent block. PMMA exhibits a glassy
transition at T g =399 K and a sub- T g (‘b’) transition at
T b =226 K. At room temperature, it is a hard, brittle
material.
The stainless steel used was the 316L SS (Z2CND17-
12). It is a high purity steel as regards inclusions and
very low in carbon. It was supplied rapid-quenched after
hot rolling. Its chemical composition is shown in Table
1. It is in conformity with the standards for manufacture
of orthopaedic implants (ISO standard 5832-1).
The metal samples were 10 mm´10 mm´20 mm par-
allelepipeds. The friction surface was first polished with
diamond paste to 1
Ò
m.
Before each test, the metal and polymer samples were
cleaned for 10 min in ethyl alcohol in an ultrasonic cle-
aner and then dried in air. The mechanical characteristics
of the materials are shown in Table 2.
The fretting tests were conducted on a TRIBOM-
INES
m
m to a roughness of Ra
=
0.03
m
device installed on a SCHENCK-PSA 10 servo-
hydraulic traction–compression machine. The cylindrical
sample of PMMA was mounted on a fixed component
provided with a conical aperture through which the con-
tact surface can be observed. The plane sample of 316L
SS was fixed in a ball joint, enabling the parallelism of
the surfaces in contact to be ensured. The length of the
cylinder/plane contact was 15 mm. Once adjusted, the
ball joint was fixed immobile relative to the hydraulic
cylinder by a clamping screw. The whole experimental
set-up is shown in Fig. 2.
one cylindrical surface of radius R
Ò
4970082.007.png
B. Tritschler et al. / Tribology International 32 (1999) 587–596
589
Table 1
Chemical composition of 316L SS stainless steel (% w/w)
Cr
Ni
Mo
C
Si
Mn
P
S
Fe
17.3
13.7
2.68
0.029
0.63
1.70
0.025
, 0.01
Bal.
Table 2
Mechanical properties of the materials used (suppliers’ data)
Poisson’s ratio n
Young’s modulus E
Yield stress R c
Ultimate tensile strength
(GPa)
(MPa)
(MPa)
PMMA
0.35
3.5
30
70
316L SS
0.3
200
280
640
Fig. 2. TribominesÒ
fretting set-up.
m. The selected cyclical frequency
of 1 Hz is comparable to the average walking frequency
of the human body. The shape of the cycles is sinusoidal.
The movements were monitored in the immediate vicin-
ity of the contact by means of inductance sensors situ-
ated on either side of the device. A pre-stressed piezo-
electric load-cell was installed on the fixed component
in order to measure the tangential force F T . The contact
interface between the two samples was displayed on a
±
40 to
100
m
monitor screen via a CCD camera linked to an optical
system. The metal sample was connected to a TACUS-
SEL PGP 201 potentiostat and a saturated calomel refer-
ence electrode (SCE) was placed close to the contact in
order to monitor changes in corrosion-free potential dur-
ing the test. A system of data acquisition and processing
was set up to study changes in the coefficient of friction,
to record the F T –displacement cycles (or F T D cycles),
and to make electrochemical measurements throughout
the test. It must be noted that the shape of the fretting
cycles obtained was, for all the tests conducted, charac-
teristic of a gross slip regime within the contact. The
fretting tests were performed in a Ringer’s solution in
The normal load F N , applied using calibrated masses,
lay within the range 0.7–6 N/mm, in order to remain
in the elastic domain of the polymer. The displacement
amplitude imposed on the metal sample lay within the
range
±
4970082.001.png
590
B. Tritschler et al. / Tribology International 32 (1999) 587–596
order to simulate human physiological fluids. This sol-
ution was prepared in the laboratory to the following
formula: NaCl to 8.5 g/l, KCl to 0.25 g/l, CaCl 2 to 0.22
g/l, NaHCO 3 to 0.15 g/l, with pH value 7.8±0.1. The
volume of material removed during the test was evalu-
ated by means of a profilometric technique. After the
test, the two metal and polymer samples were carefully
cleaned in ethanol in an ultrasonic cleaner in order to
remove all traces of debris from the surface. After dry-
ing, a TALYSURF profilometer took the profiles at reg-
ular intervals across the entire friction surface. The vol-
ume of material worn away was then calculated by
summing all the elementary volumes thus measured.
Calculation by this method provides a more accurate
assessment of the volume of abrasion than the weighing
method, which would only yield an approximate value
due to the small quantities of material worn off by fret-
ting.
The cell containing the stationary Ringer’s solution
was left open at room temperature in the laboratory
atmosphere. Once the samples were placed in the sol-
ution and after application of the load, a 30 min wait
ensued before the fretting test was started in order to
allow the free potential to stabilise.
dispersion X-ray analyses reveal mainly iron and chro-
mium oxides. The surrounding reddish deposit consists
primarily of Fe 2 O 3 iron oxides.
The PMMA shows an accumulation of fine debris at
the limits of the contact, on both sides of the friction
zone (Fig. 3). Orange-coloured deposits similar to those
observed on the 316L SS sample are also seen all around
the friction surface, although in smaller quantities. These
are the corrosion products resulting from the electro-
chemical process occurring on the surface of the metal
and little by little transferred onto the surface of the
PMMA during the test.
3.2. Evolution in coefficient of friction and free
corrosion potential
Four stages were observed in coefficient of friction or
free corrosion potential changes during the fretting tests.
Fig. 5 shows a typical example of the curves obtained
for the tests performed in Ringer’s solution. For more
clarity, the evolution of both coefficient of friction and
free potential were plotted using a logarithmic abscissa.
The first stage, which is very rapid, corresponds to a
reduction in the coefficient of friction during the early
fretting cycles, associated with a drop in free potential.
This initial fall is followed by a slower decrease of the
coefficient of friction, while the corrosion potential rises
again slightly. The second stage corresponds to an
increase in the coefficient of friction associated with a
decrease of free potential, though this decrease is less
abrupt than that observed at the beginning.
The first few cycles of the test cause the collapse of
the layer of oxide initially formed on the surface of the
metal, after which strong interaction is observed between
the two surfaces now laid bare. This marks the beginning
of the damage to the surface of the PMMA and the cre-
ation of the first debris which will cover the surface of
the metal. Those particles act as a steady-state lubricant
film between the counterfaces since the tangential force
is applied. This pseudo repassivation causes a new rise
in potential with a drop in the coefficient of friction due
to the accommodation of the movement by this new
3. Experimental results
3.1. Observation and analyses of the surfaces
The fretting tests carried out in Ringer’s solution led
to considerable damage to both surfaces. Fig. 3 shows
the appearance of the friction zone on the samples,
observed using a JEOL-840 scanning electron micro-
scope. In both cases, numerous parallel scratches in the
direction of friction are observed. Visual examination of
the friction zone on the 316L SS shows deep traces of
wear, shiny and scratched in appearance. Around the
edge of this zone, a wide orange-red deposit is observed,
revealing considerable corrosion phenomena. Fig. 4
shows debris resulting from wear located at the edge of
the friction zone at the surface of the 316L SS. Energy
Fig. 3. SEM observation of wear surfaces (frequency 1 Hz, displacement
±40
mm, normal force 4 N/mm, duration of test 20,000 cycles, Ringer’s
solution). (a) View of surface of 316L SS. (b) View of surface of PMMA.
4970082.002.png
B. Tritschler et al. / Tribology International 32 (1999) 587–596
591
m, normal force 4 N/mm, duration
of test 20,000 cycles, Ringer’s solution). Right: EDX analysis of debris; presence of oxides of iron and chromium.
±
40
m
Fig. 5. Coefficient of friction and free corrosion potential as a function of number of cycles. Contact PMMA Ø20/316L SS plane in Ringer’s
solution. Frequency 1 Hz, displacement
40
m
freshly formed third body. However, the cleansing action
of the surrounding aqueous medium causes the debris to
be carried away more or less rapidly to the edge of the
friction area. A flow of debris, gradually carried away
from the contact surface, builds up. The common dam-
age to the two surfaces results in a regular decrease in
corrosion potential to values which are more negative
(less “noble”)—showing increased electrochemical
activity of the stainless steel surface—together with an
increase in the coefficient of friction. It is pertinent to
remark here that during the first few hundred cycles, any
variation in the coefficient of friction is immediately fol-
lowed by a change in free potential, irrespective of the
operating conditions applied. This linked progression of
both values has already been shown in previous stud-
ies [22].
By in situ observation it is possible to relate the differ-
ent values measured to the phenomena of deterioration
of the polymer/metal interface. Damage to the surfaces
is visible from the very first cycles. The consequence of
the formation of PMMA debris is the appearance of a
white band which adheres to the surface of the metal
before gradually disappearing. During the period of rise
of the coefficient of friction, increasing deterioration of
the contact interface is observed, revealed by the appear-
ance of scars parallel to the direction of movement. An
increasing presence of fine, snowy debris at the edge of
the contact zone may also be observed.
The third stage of development corresponds to a slow
decrease in the coefficient of friction. During this period,
the presence in situ of increasingly marked scarring is
seen on the surface of the 316L SS, together with the
Fig. 4. SEM observation of wear debris on the surface of the 316L SS (frequency 1 Hz, displacement
m, normal force 4 N/mm. 1, breakdown of the oxide layer initially formed and pseudo repassivation;
2, flow of debris—deterioration of the contact interface; 3, degradation of the stainless steel surface; 4, equilibrium in the tribosystem.
±
4970082.003.png
Zgłoś jeśli naruszono regulamin